利用非接触测量技术研究颈椎活动度
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:

国家自然科学基金资助项目(11172161),上海市科委基金(11195820900),上海市教委基金(12ZZ092),上海市重点学科建设项目(S30106)


Evaluating range of motion in cervical spine by the non-contact measurement technology
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的 颈椎的转动范围较大,传统的数字图像相关(digital image correlation, DIC)算法不能准确测量颈椎的活动度。本文提出一种改进的DIC算法,可以实现大转角的变形测量。方法 采用一种子区窗口旋转的相关匹配法,通过引入子窗口的旋转角作为一个新的迭代量,和传统的位置参数(x,y)一起进行Newton-Rapshon迭代运算。在初始时只需提供位置和角度的迭代初值,即可求出精确的位移和角度值,同时通过平动实验和转动实验验证该方法的精度和可靠性。结果 平动实验证实,本文方法与传统的DIC具有相同的效果,位移测量精度在0.5%以内;转动实验证实,本文方法可以测量任意转角的变形,角度误差与旋转量无关,角度误差在0.5°以内。将该方法用于颈椎的压缩实验,准确地测量了颈椎在压缩过程中的活动度。结论 对比传统的DIC,子区窗口旋转的相关匹配法实现了颈椎在各种载荷作用下大范围活动度的测量,为颈椎的生理稳定和生理活动的测量判断提供了有效的方法。

    Abstract:

    Objective To develop an improved DIC (digital image correlation) algorithm suitable for measuring large ROM (range of motion) of the cervical spine, as traditional DIC algorithm is not capable of accurately measuring ROM of the cervical spine due to its large rotation angles. Methods An algorithm which allowed rotation of the subset window was proposed to achieve robust correlation matching in the measurement. A new iterative variable, which represented the orientation of the subset, was introduced and incorporated in the Newton-Raphson iteration method together with the position variables (x,y). By assigning an initial guess to these variables individually, the precise location and rotation angle could be determined in the deformed image. The precision of the proposed method was evaluated by translation and rotation experiments. ResultsThe translation experiment confirmed that the proposed method had the same accuracy as the traditional DIC, and the displacement measurement accuracy was within 0.5%. While the rotation experiment indicated that the proposed method could measure the deformation at any angles with precision less than 0.5°. The method was then applied to the measurement of ROM of cervical spine subjected to compressive loads and received good results. Conclusions Compared with the traditional DIC algorithm, the proposed method can achieve accurate measurement with large ROM for cervical spine tests with different loads, and provide an effective means for assessing the stability and physiological activities of cervical spine.

    参考文献
    相似文献
    引证文献
引用本文

侯义荣,李凯,东人,张东升.利用非接触测量技术研究颈椎活动度[J].医用生物力学,2013,28(1):97-102

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2012-05-21
  • 最后修改日期:2012-04-26
  • 录用日期:
  • 在线发布日期:
  • 出版日期:
关闭