自主研制胸锁关节解剖锁定钢板的生物力学对比
DOI:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:

四川省科技计划项目(2014JY0248),泸州市科技计划项目[2016-S-67(14/23)],西南医科大学基金项目(2013ZRQN069),西南医科大学附属医院基金项目(2017-PT-48)


Biomechanical Comparison of Self-Developed Anatomical Locking Plate for Sternoclavicular Joint
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的 将自主研制的胸锁关节解剖锁定钢板与桡骨远端斜T型锁定钢板及胸锁钩钢板进行生物力学对比分析,观察其生物力学方面的特点及优势。方法 选择9具防腐、湿润成人尸体,其中男性6具,女性3具。完整取下胸锁关节标本造成胸锁关节全脱位模型。将9具标本模型的两侧胸锁关节随机编号并配对分组:A组(实验组,胸锁关节解剖锁定钢板)、B组(对照组1,桡骨远端斜T型锁定钢板)和C组(对照组2,胸锁钩钢板),每组6个胸锁关节。将所有标本两侧安置钢板后固定于万能力学试验机上,分别进行锁骨远端负载、锁骨远端扭转及胸骨柄部螺钉抗拔出3项生物力学实验。结果 锁骨远端负载实验:3组标本载荷量位移均呈线性关系,实验组的抗压缩形变性能强于两个对照组。锁骨远端扭转实验:扭角与扭矩之间均呈线性关系,实验组的抗扭转形变性能强于两个对照组。胸骨柄部螺钉抗拔出实验:A、B、C组胸骨柄部螺钉最大抗拔出力比较,差异有统计学意义(P<0.05)。实验组胸骨柄部螺钉抗拔出性能强于两个对照组。结论 自主研制胸锁关节解剖锁定钢板在抗压缩、抗扭转、胸骨柄部螺钉抗拔出等生物力学性能方面显著优于斜T型锁定钢板及胸锁钩钢板,能够为临床上治疗胸锁关节骨折脱位提供一款较为理想的内固定器械。

    Abstract:

    Objective To observe biomechanical characteristics and advantages of the self-developed anatomical locking plate of sternoclavicular joint by comparison with the radial distal oblique T-shaped locking plate and sternoclavicular hook plate. Methods Nine embalming and moistening adult corpses were selected, including 6 males and 3 females. Model of complete dislocation of the sternoclavicular joint caused by complete removal of the sternoclavicular joint specimen. The bilateral sternoclavicular joints of 9 specimen models were randomly numbered, matched and divided into Group A(experimental group, sternoclavicular joint anatomical locking plate), Group B(control group 1, the radial distal oblique T-shaped locking plate) and Group C(control group 2, sternoclavicular hook plate), with 6 sternoclavicular joints in each group. All specimens were placed with steel plates on both sides and fixed on the universal mechanical test machine. Three biomechanical experiments were carried out, including loading of distal clavicle, torsion of distal clavicle and anti-pull out of sternum handle screw. Results Distal clavicle loading test: the load-displacement of specimens in three groups showed a linear relationship. The compressive deformation resistance in experimental group was stronger than that in two control groups. Distal clavicle torsion test: the relationship between torque and torsion angle was linear. The torsional deformation resistance in experimental group was stronger than that in two control groups. Anti-pullout test of sternum handle screw: there was a significant difference in the maximum anti-pullout force of sternum handle screw among the three groups (P<0.05). The anti-pullout performance of the sternum handle screw in the experimental group was better than that in the two control groups. Conclusions The self-developed sternoclavicular joint anatomical locking plate is superior to the oblique T-shaped locking plate of distal radius and the plate of sternoclavicular hook in terms of anti-compression, anti-torsion and anti-pullout of sternoclavicular screw, so as to provide an ideal internal fixation device for the treatment of fracture and dislocation of sternoclavicular joint.

    参考文献
    相似文献
    引证文献
引用本文

周举,杨琨,孙远林,向飞帆,刘煊文,阳运康.自主研制胸锁关节解剖锁定钢板的生物力学对比[J].医用生物力学,2020,35(5):608-614

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2019-10-14
  • 最后修改日期:2019-12-29
  • 录用日期:
  • 在线发布日期: 2020-10-23
  • 出版日期: 2020-10-01
关闭