Design Optimization, Function Analysis and in vitro Experiments of an Electric Stapler for Minimally Invasive Surgery
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的 开发一款新型电动吻合器,以解决现有产品转角不足、操作不便、难以掌握压榨强度等问题。方法 设计一款电动吻合器并加工出样机,利用三坐标影像仪搭建样机功能测试平台测量出样机的运动轨迹曲线,用拟合优度检验来评估理论曲线与测量曲线的一致程度;在前端不同弯曲角度下,吻合新鲜离体猪小肠组织,测试吻合口的吻钉成型率。结果 转弯运动与击发运动的测试曲线与理论曲线拟合优度检验理想,而压榨运动的测试曲线与理论曲线的拟合优度在转弯关节弯曲0°~30°时不理想,45°~60°时理想。性能实验中吻钉畸形率低于1.14%,表明弯曲角度对吻合效果无显著影响。结论 击发运动和转弯运动的运动学测量曲线与理论曲线一致,压榨运动曲线在不同弯曲角度时存在波动,但不影响吻合效果,电动吻合器的吻合效果符合临床要求。

    Abstract:

    Objective To develop a new type of electric stapler, so as to solve the problems of insufficient rotation angle, inconvenient operation and difficulty in controlling the pressing strength of existing products. Methods An electric stapler was designed and manufactured. The motion trajectory curve of the prototype was measured by using the three-coordinate imaging instrument to build functional test platform of the prototype, and the goodness of fit was used to evaluate consistency between the theoretical curve and the measured curve. The small intestine tissues of fresh pig were anastomosed at different bending angles of the front end, and the forming rate of the anastomotic stoma was measured. Results The goodness of fit between the test curve and the theoretical curve for both turning motion and shooting motion was ideal, while the goodness of fit between the test curve and the theoretical curve for pressing motion was not ideal when the turning joint was bent at 0°-30°, and was ideal when it was bent at 45°-60°. In performance test, the deformity rate of the nail was smaller than 1.14%, indicating that the bending angle had no significant impacts on the anastomotic effect. Conclusions The kinematics curves of shooting motion and turning motion are consistent with the theoretical curves. The pressing motion curves fluctuate at different bending angles, which will not affect the anastomotic effect, and the effect of the electric stapler meets the clinical requirements.

    参考文献
    相似文献
    引证文献
引用本文

葛文明,王佩瑶,刘晨旭,凌宇秀,潘贞,胡钟欣,周宇,宋成利.微创电动吻合器的优化设计、功能分析与离体实验[J].医用生物力学,2020,35(5):574-580

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2019-09-29
  • 最后修改日期:2019-11-26
  • 录用日期:
  • 在线发布日期: 2020-10-23
  • 出版日期: 2020-10-01
关闭